The role of low-temperature plasma hydrogen peroxide transfer window


Release time:

14 Sep,2022

Although there is no doubt about the role of the sterilization of the low-temperature plasma hydrogen peroxide transfer window, in order to prevent frequent abnormalities due to errors in actual operation and endanger normal start-up, it is necessary to master the appropriate and practical operation of the low-temperature plasma hydrogen peroxide sterilizer. Method.

1.Before starting, check the appearance of the low-temperature plasma hydrogen peroxide transfer window and whether the power plug is abnormal. If there is an abnormality, you must first find out the cause before starting.
2.Before starting, you need to check whether the negative pressure pump oil pump is in the conventional liquid level line, whether there is a problem with the oil color, whether there is an oil pump in the air filter at the exhaust port, etc. ; If the said situation occurs, regardless of whether the oil pump is higher than or lower than the usual normal liquid level line, regardless of the oil color and turbidity, the oil pump needs to be removed and replaced, and the oil mist separator must be removed or replaced.
3.If there is water or other evaporative chemicals entering the inner cavity of the pump, the ballast valve can be opened to carry out purification treatment; check whether there is water or smudges in the air inlet water-gas separator, and it can be cleaned up or replaced as necessary.
4.When replacing the oil pump, first operate the mechanical pump for three minutes up and down and then add some oil. Then introduce 100 ml up and down at the oil inlet to clean the oil pump, and then fill the oil pump after the waste oil is exhausted; generally, it is filled to the middle of one-half and two-thirds of the oil mirror.
5.If the low-temperature plasma hydrogen peroxide sterilizer has not been used for more than three days, be sure to carry out system inspection and treatment; and remove the protective plate on the side of the rotating shaft, whether the leaves of the rotating shaft can be rotated by hand, and how to use the appropriate energy is difficult to rotate, then mechanical pump maintenance is required.
6.Before disinfection and sterilization, check whether the door sealing ring of the low-temperature plasma hydrogen peroxide sterilizer is cleaned and deformed; check whether the filter at the suction outlet of the sterilization room is blocked; check whether there are oil stains and residues in the inner cavity of the room; check whether the low-temperature plasma transmitter is clean and dry.Check whether there is too little hydrogen peroxide, which is not enough for one sterilization application.

The hydrogen peroxide transfer window is the air purification system used in the purification room.It is suitable for short-term disinfection and sterilization of small and medium-sized bank credit facilities, equipment, objects and air in different levels of purification rooms and non-clean areas, and reasonably reduces the frequency of opening the purification room.

The hydrogen peroxide transfer window is integrated by a gasification hydrogen peroxide generator. The gasification hydrogen peroxide generator gives hydrogen peroxide vapor to the internal structure of the hydrogen peroxide transfer window, and microbial cleaning is carried out on the surface of the raw materials to avoid pollution of the raw materials into key areas of Class A and B.It is suitable for all kinds of clean and dry items that must be transported during the production process of aseptic testing, including packaging products, equipment, raw materials, parts and components, environmental monitoring instruments, etc.Enter the Class A and Class b heavily polluted areas.